Sign In | Join Free | My lightneasy.org
China Shandong Haiying Wuzhou Instrument Co., Ltd. logo
Shandong Haiying Wuzhou Instrument Co., Ltd.
Shandong Haiying Wuzhou Instrument Co., Ltd.
Active Member

5 Years

Home > BSC Biosafety Cabinet >

ISO 5 Ductless Biosafety Cabinet Chemical Reserch Filtration

Shandong Haiying Wuzhou Instrument Co., Ltd.
Contact Now

ISO 5 Ductless Biosafety Cabinet Chemical Reserch Filtration

Brand Name : HAIYING

Model Number : BSC-1300B2

Certification : CE

Place of Origin : CHINA

MOQ : 1 set

Price : Negotiable

Payment Terms : L/C, D/A, D/P, T/T, Western Union, MoneyGram

Packaging Details : Wooden case

Type : Blood Testing Equipments

Name : BSC Biosafety Cabinet

Overall size : 1500×800×2120mm

Purification area size : 1300×700×600mm

Product name : Biosafety cabinet

Model : BSC-1300B2

Downflow velocity : 0.31±0.025m/s

Inflow velocity : 0.48±0.025m/s

Contact Now

BSC Biosafety Cabinet Features:

1. High-quality material front window.
2. Timing function.
3. HEPA filter life and UV life indicator.
4. Automatic wind speed with filter is adjustable.
5. With power-off memory function.
6. Audiovisual alarm (abnormal airflow speed; filter replacement; front window is at an unsafe height.)
7. The working area is surrounded by negative pressure, which can ensure the maximum safety of the working area.
8. Audiovisual alarm (filter replacement, high-altitude windows, abnormal airflow speed, etc.)
9. Remote control: all functions can be realized, making the operation easier.
10. Interlocking function; UV lamp and front window; UV lamp and blower, LED lamp; Blower and front window.

Model BSC-1300B2
Clean Grade HEPA:ISO 5 calss ( Class100 )
ULPA:ISO 4 Class ( Class10 )
HEPA Filter HEPA:≥99.995% , @0.3μm
ULPA:≥99.999% , @0.12μm
Air tightness 99.99≤E≤99.999
Overall Dimensions 1500×800×2120mm
Working Space 1300×700×600
mm
Downflow Velocity 0.35m/s
Forward intake wind speed 0.25m/s±0.025m/s

Principles for choosing BSC Biosafety Cabinet in biosafety laboratories:

When the laboratory level is level 1, there is generally no need to use a biological safety cabinet or a level I biological safety cabinet. When the laboratory level is level two, when microbial aerosols or spills may occur, a level I biological safety cabinet can be used; when handling infectious materials, a level II biological safety cabinet with partial or full ventilation should be used ; If it involves handling chemical carcinogens, radioactive substances and volatile solvents, only Class II-B full exhaust (B2 type) biological safety cabinets can be used. When the laboratory level is level three, a level II or level III biological safety cabinet should be used; all operations involving infectious materials should use a fully exhausted type II-B level (B2 type) or level III biological safety cabinet. When the laboratory level is level four, a level III full exhaust biological safety cabinet should be used. When personnel wear positive pressure protective clothing, a Class II-B biological safety cabinet can be used.

Installation of BSC Biosafety Cabinet:

1. The biological safety cabinet shall not be placed on its side, impacted, or collided during transportation, and shall not be directly attacked by rain, snow, or sun exposure.

2. The working environment of the biological safety cabinet is 10~30℃ and the relative humidity is <75%.

3. The equipment should be installed on a horizontal surface that cannot be moved.

4. The equipment must be installed close to a fixed power socket. When the external exhaust system is not connected, the top of the device must be at least 200mm away from the obstacle at the top of the room, and the distance between the back and the wall must be at least 300mm to facilitate the smooth flow of external exhaust. Maintenance of safety cabinets.

5. In order to prevent airflow interference, it is required not to install the equipment in the passage of people, and not to make the operating port of the sliding front window of the biological safety cabinet face the door and window of the laboratory or be too close to the door and window. Places that may disturb the airflow.

6. When used in high altitude areas, the wind speed must be recalibrated after installation.

ISO 5 Ductless Biosafety Cabinet Chemical Reserch Filtration


Product Tags:

B2 Ductless Biosafety Cabinet

      

B2 ISO 5 biosafety cabinets

      

BSC Ductless Biosafety Cabinet

      
Buy cheap ISO 5 Ductless Biosafety Cabinet Chemical Reserch Filtration product

ISO 5 Ductless Biosafety Cabinet Chemical Reserch Filtration Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: Shandong Haiying Wuzhou Instrument Co., Ltd.
*Subject:
*Message:
Characters Remaining: (0/3000)